Edwards APG100 series Active Pirani vacuum gaugesare available in 2 models. The \nAPG100-XM is the standard model and measures to 10-3 mbar. The APG100-XLC is a \ncorrosion resistant version with measurement to 10-4 mbar. \n \nBoth Edwards APG100 gauges feature compact size for easy installation, a linear output, and a replaceable \nsensor tube. The new gauges are compatible with all Edwards Vacuum TIC instrument \ncontrollers and other active gauge controllers and displays. They are also CSA, C/US \napproved as well as fully RoHS compliant due to their lead-free construction

Features:


  • Cable connections and gauge adjustment conveniently located, thereby minimizing the space envelope required for access
  • Sensor tube can be baked to 150 ºC
  • Adjustable set-point for simple process control and interlocking
  • Remote calibration possible
  • Linear output - 1 volt per decade for easy interface with vacuum control systems
  • NW16 and NW25 flange options for easy connection to vacuum systems – other flange options on request
  • LED status indicator shows normal and fault conditions

Technical Data:


Measurement range
  • APG100-XM Atmosphere to 10-3 mbar
  • APG100-XLC Atmosphere to 10-4 mbar
Accuracy
  • APG100-XM Typically +/- 15% at <100 mbar
  • APGX100-XLC Typically +/- 15% at <10 mbar
Maximum over-pressure 10 bar absolute
Temperature range
  • Operating 5º to 60º C
  • Storage -30º to 70º C
  • Bake-out with electronics removed 150 ºC
  • Humidity 80% RH up to 31 ºC decreasing
  • linearly to 50% RH at 40 ºC and above
  • Maximum altitude 3000 m
  • Filament temperature 100 ºC above ambient
Electrical data
  • Electrical supply voltage 15 to 30 V d.c. nominal
  • 13.5 V d.c. minimum
  • 32 V d.c maximum
  • Power consumption 1 W
  • Output signal 0 to 10 V d.c. nominal
  • Set-point – open collector transistor
  • Rating 30 V d.c. 100 mA
  • Range of set-point 1.8 to 9.2 V dc
  • Fixed hysteresis 500 mV (1/2 decade)
  • Level setting resolution 6 mV
Authenticity Icon 100% Original and Quality
Guarantee Icon 24 Months Guarantee

Recently Viewed

Choose Ezzi Vision

Vacuum, Lithography, Thin-Film Experts

Experience top-tier repair and maintenance services for glovebox and vacuum equipment, backed by industry expertise across Australia.

Contact Us

Product Compare

Comparing

My Wishlist

Love it? Add to your wishlist

Your favorites, all in one place. Shop quickly and easily with the wishlist feature!